Recall Notice: Roche ACCU-CHEK FlexLink Plus Infusion Set: Class I Recall - Potential for Under-Delivery of Insulin

A Class I Recall has been issued for Roche ACCU-CHEK FlexLink Plus Infusion Sets. There is the potential for under-delivery of insulin due to a tube (cannula) which may become kinked or bent when inserting the ACCU-CHEK FlexLink Plus infusion set. If gone unnoticed, this can result in the under-delivery or no delivery of insulin. This can lead to elevated blood glucose levels (hyperglycemia). Hyperglycemia can lead to many serious health complications including death. Of important note, this recall only applies to the ACCU-CHEK FlexLink Plus infusion sets that were launched in November 2010. ACCU-CHEK Ultraflex, other Accu-Chek infusion sets or insulin pumps are not affected by this recall and can be continued as directed by a physician or other qualified healthcare provider. Roche requests that its consumers stop using the ACCU-CHEK FlexLink Plus infusion sets and return the unused products. Patients are encouraged to contact their health care providers or caregivers to determine if changes to their therapy are needed and how to temporarily continue insulin pump therapy without the ACCU-CHEK FlexLink Plus infusion set. Click here for more information.